Remote Control
R&S
®
RTM20xx
417User Manual 1317.4726.02 ─ 01
Return values:
<ByteCount> Number of words (bytes)
Usage: Query only
BUS<b>:CAN:FRAMe<n>:BYTE<o>:STATe?
Returns the state of the specified data byte.
Suffix:
<b>
.
1..4
Selects the bus.
<n> *
Selects the frame (1...n).
<o> *
Selects the byte number (1...n).
Return values:
<ByteStatus> OK | UNDF
UNDF: Undefined
Usage: Query only
BUS<b>:CAN:FRAMe<n>:BYTE<o>:VALue?
Returns the decimal value of the specified byte.
Suffix:
<b>
.
1..4
Selects the bus.
<n> *
Selects the frame (1...n).
<o> *
Selects the byte number (1...n).
Return values:
<ByteValue> Decimal value
Usage: Query only
CAN - Search
SEARch:PROTocol:CAN:CONDition................................................................................418
SEARch:PROTocol:CAN:FRAMe.....................................................................................418
SEARch:PROTocol:CAN:ACKerror..................................................................................419
SEARch:PROTocol:CAN:BITSterror.................................................................................419
SEARch:PROTocol:CAN:CRCerror..................................................................................419
SEARch:PROTocol:CAN:FORMerror...............................................................................420
SEARch:PROTocol:CAN:FTYPe.....................................................................................420
SEARch:PROTocol:CAN:ITYPe.......................................................................................420