20
Calibration and Linearization
During calibration, the balance is adapted to changes in ambient conditions.
You should recalibrate your balance each lime you set it up in a different area or
when the ambient conditions change (fore example, temperature or barometric
pressure). When these conditions are constant, the balance should be calibrated
once a day. To meet the highest requirements for accurate weighing, we recom-
mend that you calibrate the balance before each weighing series.
Make sure that the draft shield is closed before calibration or linearization Starts.
You can abort any calibration or linearization procedure by pressing the
key (28),
The balance offers you various calibration and linearization functions. The func-
tion you select is indicated in the display by one of the following special codes;
»CALI« The balance has built-in calibration weights and can be calibrated
with the
key (“Quick-CAL" - see page 22)
»CALI« The calibration or linearization function is activated
»C.I.« Internal calibration
»C.t.« Calibration test
»L.I.« Linearisieren intern
»C.E.« Eternal calibration