EasyManuals Logo

Sartorius mc1 User Manual

Sartorius mc1
136 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #14 background imageLoading...
Page #14 background image
13
Important Note
lf the
symbol flashes, this means that the balance wants to self-calibrate. You
do not need to interrupt your weighing procedure, because the balance will wait
until it detects that you have stopped weighing for one complete minute.
Afterwards, the balance will perform fully automatic infernal calibration. The
symbol will flash until the balance Starts this fully automatic procedure on its own
or until you activate one of the calibration functions by pressing a key (see p p.
22).
Turning the Display On and Off (Standby Mode)
Press the
key (27) to turn the display on or off.
Self-Test
After the balance has been turned on, an automatic self-test of the balance's elec-
tronic circuitry is performed, and the draft shield closes automatically.
It ends with a readout of "0.00000 g" or "0.0000 g" (depending on your bal-
ance model or the display mode selected).
Important Note for Automatic Operation with a Robot
To be on the safe side for automatic Operation of the -balance with a robot, you
should turn off the fully automatic draft shield function (see "Part III: Balance Op-
erating Menu"). Otherwise, if a power outage occurs and the balance is auto-
matically restarted, the robot arm may accidentally hit the draft shield as it is clos-
ing.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius mc1 and is the answer not in the manual?

Sartorius mc1 Specifications

General IconGeneral
BrandSartorius
Modelmc1
CategoryScales
LanguageEnglish

Related product manuals