Determination of the Residual Weight in Percent Code 2 1 5
Additional menu codes used in the example:
Function Code
With data ID code 7 2 2
As an alternative, you can select
this Output parameter:
Ref. % and weight 7 1 2 Reference weight 7 1 3
Application: quick determination of the dry weight of a sample
Step/(key lD)/key Readout Data output/printout
Place Container on the pan + 20.09905 g
,
0.00000 g
Add prepared sample
to the container
+ 4.61625 g
Store ref. wt. as 100%:
STO REF
+ 100.0 %
pRef
Wxx%
+ 100 %
+ 4..61625 g
Remove container with
sample; dry sample
xx.x %
Reweigh sample in the
container
+ 72.5 %
Net
+ 72.5 % Prc + 72.5 %
Toggle to residual
weight readout:
+ 3.34678 g
+ 3.34678 g N + 3.34678 g
The data ID codes mean
pRef
Wxx%
Prc
N
+
+
+
+
100 %
4.61625 g
72.5 %
3.34678 g
reference percentage
net reference weight xx%
calculated percentage
net residual weight