Specifications
Model RC 210S RC 210P RC 210D RC 250S
Weighing range structure SuperRange PolyRange DualRange SuperRange
Capacity/weighing range levels g 210 60 / 110 / 210 52 / 210 250
Readability mg 0.01 0.01 / 0.02 / 0.05 0.01 / 0.1 0.1
Tare range (by subtraction) g -210 -210 -210 -250
Reproducibility (standard deviation)*
for weighing range
mg
g
≤
±0.015 ±0.02 ±0.03
0-50 50-100 100-210
≤
±0.02 /0.04/0.05
≤
±0.02 /0.1
≤
±0.1
Linearity mg
≤
±0.12
≤
±0.15
≤
±0.03/0.2
≤
±0.2
Linearity within 5 g mg
≤
±0.02
≤
±0.02
≤
±0.02
Response time (typical) s
≤
10
≤
10
≤
10
≤
2.5
Response time in the IQ-Mode (0.01 %) s 2.5 2.5 2.5 1.5
Adaption to ambient conditions and application requirements 4 optimized filter levels
Display update (depends on the filter level selected) s 0.2 – 0.4 (selectable)
Allowable ambient temperature range for operation
o
C +5 … +40
Operating temperature range
o
C +10 … +30
Sensitivity drift within + 10 … + 30
o
C /
o
C
≤
±1 ·
10-6
Moisture-proof rating acc. to DIN 40040 Class F, non-condensing
Pan size mm Ø 90 / 3.5”
Weighing chamber (Ø x H), approx. mm 166 x 248/ 6.5 x 9.8”
Net weight, approx. kg 12 26.4 lbs
AC power source/power requirements V~ power supply. 230 or 115. +15 % . . . –20 %
Frequency Hz 48 – 60
Power consumption incl. power supply, approx. VA max.: 28; average: 16
Selectable weight units g, kg, ct, lb, oz, ozt, tlh, tls, tlt, gr, dwt and mg
Applications programs
mass unit conversion, tare memory, net total, weighing in percent,
over/under checkweighing (classification and filling), counting
Automatic zero tracking function (can be turned off by menu code) standard feature
Built-in interface
RS 232 C-S/V24-V28, RS 423/V10; 7-bit; parity;: even, mark, odd, space;
transmission rates: 150 … 19,200 baud; 1 or 2 stop bits; software/hardware handshake
Standard features/equipment supplied
Dust cover x x x x
Power supply x x x x
Attachment for below-balance weighing x x x x
Interior weighing chamber draft shield x x x x
Built-in automatic calibration available on request
*
= standard deviation of the reproducibility according to DIN 1319, Part 3
3