13
Checking Net Weights-Weight Readout Code 2 2 3
To store a weight readout as the target weight, press
.
The target weight is shown in the analog display as a bar graph. The last segment of this bar graph ends exactly below the
equal sign of the checkweighing indication.
lf the weight of a product is "okay," the last segment of the bar graph will be within the tolerance range of the checkweighing
indication. You can also read the actual weight in digits.
Additional menu codes used in the example:
Function Code
With data ID code 7 2 2
As an alternative, you can select
this output parameter.
Tolerance + target 7 1 2 Target weight only 7 1 3
Application: weighing in the same amounts of sample
Step/(key ID)/key Readout Data output/printout
, place sample container on
the pan,
0.00000 g
Weigh in required sample
amount
+ 40.10280 g
Store: (STO)
1.0 % Lim + 1.0 %
Setp + 40.10280 g
+ 40.10280 g
Remove Container + ideal
sample; place another
Container on the pan;
0.00000 g
Weigh in sample using the
analog display as a guide
Checkweighing
indication
+ 40.15295 g
Continue as directed above
+ 40.15295 g
N
+ 40.15295 g
The data ID codes mean
Lim
Setp
N
+
+
+
1.0 %
40.10280 g
40.15295 g
over/under limits in percent
target weight stored
net sample weight