Page #23 background imageLoading...
Page #23 background image
22
INTERNAL CALIBRATION
Quick-CAL Function using the
key:
Unload the pan, close the draft shield and tare.
When the balance displays a zero readout, press the
key (20).
"C" will now be displayed. The built-in calibration weights are internal~ applied
by servomotor and removed at the end of calibration.
lf external interference affects the calibration procedure, you may obtain a brief
display of the error message "Err 02."
In this case, tare and then press the
key again when a zero readout appears.
An acoustic Signal indicates the end of calibration.
Internal Calibration Using the Tare Control:
Calibrate the balance using the tare control if an application program (such as the
tare memory) is assigned to the
key by menu code in the balance operating
menu (see Parts III and IV).
Press the tare control (24) for at least 2 seconds until "C.l." and "CAL” are dis-
played (next to the
key).
Unload the pan, make sure the draft shield is closed, and tare. When the balance
displays a zero readout, press the
key (20).
“C" will now be displayed. The built-in calibration weights are internal by applied
by servomotor and removed at the end of calibration.
If external interference affects the calibration procedure, you may obtain a brief
display of the error message "Err 02."
In this case, tare and then press the
key again when a zero readout appears.
An acoustic signal indicates the end of calibration.