21
Fully Automatic Calibration
lf the
symbol flashes, this means that the balance wants to self--calibrate. You
do not need 1o interrupt your weighing procedure, because the balance will wait
until it detects that you have stopped weighing for one complete minute.
Afterwards, the balance will perform fully automatic internal calibration.
During this procedure, the draft shield should be closed 1o ensure that calibration
is done correctly.
The symbol will flash until the balance Starts this fully automatic procedure on its
own o r until you activate one of the calibration functions by pressing a key (sees
next page).
The balance can perform fully automatic internal calibration two hour after it has
been connected to line power or once the temperature has changed by ±1.5 de-
grees since the last calibration process. Independently of these factors, the bal-
ance can perform a fully automatic calibration after four hours at the latest.
To turn off the fully automatic calibration function by menu code, refer to "Part III:
Balance Operating Menu."
In the fully automatic mode, the draft shield closes after
or the tare control is
pressed.