19
Counting
A weight readout is stored as a reference sample quantity (factory setting: 10 pcs). To change this value, see the description on
page 30.
(The reference sample quantity is displayed as long as the reference sample weight has not yet been stored for counting.) While
counting (piece count readout), you can press
identified by - OPT - to Update the reference sample for optimising the
accuracy (approx. double the reference sample quantity). In the process, the reference weight is recalculated based on the total
reference sample weight and the piece count displayed.
Counting Small Parts Code 2 1 4
Additional menu codes used in the example:
Function Code
With data ID code 7 2 2
As an alternative, you can select
this output parameter:
Ref. quantity + ref. wt. 7 1 2 Reference weight 7 1 3
Application: counting bulk quantities of items with the same weight
Step/(key lD)/key Readout Data output/printout
Place Container on the pan + 20.09905 g
,
0.00000 g
Add 10 sample parts Store:
STO REF
+ 0.20019 g
+ 10 pcs
nRef
+ 10 pcs
wRef + 0.020019 g
Fill Container with desired
quantity
+ 3784 pcs
+ 3784 pcs Qnt + 3784 pcs
Toggle to weight readout:
press
+ 75.75190 g
+ 75.75190 g N + 75.75190 g
STO
1.0 % Lim + 1.0 %
To Update ref. sample, add
more parts
(approx. 2x nRef)
press: OPT
+ 20 pcs
+ 0.020014 g
wRef
+ 0.020014 g
The data ID Codes mean
nRef
wRef
Qnt
N
+
+
+
+
10 pcs
0.020019 g
3784 pcs
75.75190 g
reference sample quantity (piece count)
reference weight
calculated piece count
net weight