Troubleshooting Guide
Problem … Causes Solution
No segments appear in the weight
display (19)
No line power available
The power supply is not plugged in
Check power supply
Plug in the power supply
The display shows “Err 54” or ”L”
The weighing pan (7) is not in place
Position the pan
The display shows ”H”
The load exceeds the capacity of the
balance
Unload the balance
The display (19) briefly shows
”Err 02”
The key (20) was not pressed when the
display indicated a zero readout during
calibration
The balance is loaded
Press the tare control and re-press
the
key
Unload the balance
The display (19) briefly shows
”Err 03” or ”Err 04”
The balance is in the warmup phase
The weighing System is affected by drafts
or vibrations
After plugging the balance into line
power, allow for at least 2 hours'
warmup
Set up the balance in another area
The special code “C” does not go
out in the weight display
The balance is not ready to calibrate or is
in the warmup phase
The weighing system is affected by drafts or
vibrations
The draft shield door is not closed
The port for the below-balance weighing
hanger is open
After plugging in the balance using
the power supply, allow for at least 2
hours' warmup
Access the menu to select the correct
code for the weighing environment
(see Part III)
Check draft shield function and dose
door
Fasten the closing plate to dose the port
for below balance weighing
The special code does not go
out in the weight display (19)
None of the keys has been pressed after the
balance was turned on
Press a key
The weight readout changes
constantly
Unstable ambient conditions
Too much Vibration or the balance is
exposed to a draft
The door of the draft shield is not
completely closed
A draft shield cover is not in place
A foreign object is caught between the pan
and the bottom of the weighing chamber
The port for the below-balance weighing
hanger is open
The sample does not have a stable weight
(absorbs moisture or evaporates)
Set up the balance in another area
Access the menu to select the correct
code for the weighing environment
(see Part III)
Close the draft shield door
Position the draft shield cover
Remove foreign object
Fasten the plate to dose the port for
below-balance weighing
The weight readout is obviously
wrong
The balance is not calibrated
The balance was not tared
before weighing
The air bubble in the level indicator (10) is
not within the circle
Calibrate (see pp. 20-25)
Tare before weighing
Level the balance (see page 11)