Error Codes
These error codes enable you to identify any operator errors that occur.
Readout Meaning Reason/Solution
Err IM
The tare control is blocked when tare memory has
a value stored.
The tare functions are blocked with respect to
each other. To release the tare control, clear the
tare memory of the particular function
key (press (
).
Err II
You tried to store a value not allowed in the tare
memory.
The readout was negative, for example - check the
sample on the pan.
Err 20
The key was pressed before beginning an
application program.
For weighing in percent, a reference percentage
was not yet stored, for example.
Err 21
The menu code for the reference percentage or
reference sample quantity is blocked.
The code 2 3 1 is selected in the balance
operating menu. Change it to 2 3 2 or 2 3 3.
Err 22
Error during storage of the reference for weighing
in percent or counting.
The weight is too light or a sample is not on the
weighing pan.
Err 23
Error during storage for over/under
checkweiqhinq.
A sample is not on the weighing pan or the
readout has a negative siqn.