11
Balance Operating Parameters
How to Adapt the Balance to Ambient Conditions
To adapt your balance to ambient conditions, you may need to change the response time (see the Product Data Sheet).
Code
Very stable conditions 1 1 1
Stable conditions * 1 1 2
Unstable conditions 1 1 3
Very unstable conditions 1 1 4
Standard Weighing Mode - Manual Filling Mode
You can optimally adapt your balance to meet either of these requirements.
In the manual filling mode, the display compensates for fluctuations of the load on the balance so that you obtain a steadier
readout.
Code
Standard weighing mode * 1 2 1
Manual filling mode 1 2 2
Stability Range
The stability range in digits works together with the stability symbol (unit symbol).
When the stability symbol is displayed, the weight readout is stable within the defined range.
Readout is stable within ± Code
0,25 digit ** 1 3 1
0,5 digit 1 3 2
1 digit 1 3 3
2 digits * 1 3 4
4 digits 1 3 5
8 digits 1 3 6
* = Factory setting
** = Setting not available for balances with 0.01 mg readability