3
Insert the suction tubing (B) into the
suction filter (A).
4
Replace the solvent in reservoir bottle with water, and place the suction tubing
back into the bottle.
5
Press [CE] to display the initial screen.
6
Deliver water at 1 mL/min for 10 minutes.
7
Check for air bubbles inside the suction tubing.
Replace the suction filter if air bubbles are present.
If Ghost peaking occurs, rinse the suction filter. Refer to System Guide "5.3.2
Symptoms Presumably Due to an Error, Based on a Chromatogram".
118
LC-40D X3 CL LC-40D XR CL
4 Maintenance