5.1.5
Other Specifications
Item Specification
Liquid-Contacting Part Materials
SUS316L, PEEK, ruby, sapphire, HastelloyC,
polyethylene
Operating Temperature Range
4 °C to 35 °C
Humidity Range 20 % to 85 %
Available pH Range 1 to 14
Dimensions
W260 mm ´ H140 mm ´ D500 mm (Excluding
protruding parts)
Mass 10 kg
Power Supply
Power Supply Voltage
(indicated on the
instrument)
100 V AC to 240 V AC
* Mains supply voltage fluctuations are not to
exceed 10 % of the nominal supply voltage.
Power Consumption 150 VA
Frequency 50/60 Hz
Rated Breaking Capacity
50 A
* Connect the instrument to a power outlet
that is equipped with a circuit breaker that
shuts off the current at the described value or
less.
Installation Environment (IEC)
Installation Category , Pollution Degree 2,
Altitude 2000 m max.
Install indoors.
Plunger Rinsing Line Automatic rinsing kit as standard equipment
Pressure Limiter Operations Upper/lower limits can be set
Error Display
Exist (Error display and stop at the time of
malfunction)
128