6
Press [PURGE], and then [ENTER].
During delivery, first press [pump]
and [ENTER] to stop the delivery,
then press [purge] and [ENTER].
IF the drain valve knob is turned more than 180°, any mobile phase that drains
out may contain air bubbles. This is normal.
3.2.2
Operation in Constant Flow Delivery Mode
There are two basic operating modes; one is in constant flow delivery mode and the
other is in constant pressure delivery mode.
1
Turn the drain valve knob clockwise as far as possible to close the drain valve.
2
Press [CE].
The initial screen is displayed.
3
Press [Enter] once and set the flow
rate to be delivered.
To
set 1 mL/min, press [1] and [Enter].
4
Press [PUMP], and then [ENTER].
[PUMP] on the display is highlighted and
the pump starts operation.
5
Check that the pump outlet pressure is increasing on the pressure indication
display.
20
LC-40D X3 CL LC-40D XR CL
3 Operation