4.15.2
Decontamination cleaning
After handling the hazardous substances, using the following procedure.
1
Turn off the power of the column oven.
The gas sensor may defect volatile gas generated from organic solvent.
2
Clean with a cloth soaked in Aqueous ethanol solution (80%) or aqueous sodium
hypochlorite solution (0.05%) and wrung tightly.
Do not leave metal surfaces wet.
This can cause rusting and discoloration.
4.16
Cleaning of the Automatic Rinsing Kit
Necessary parts
Part Name
Part Type Part No. Quantity
WASHING PUMP SEAL KIT
Replacement
part
228-56201-41 1
If the 10% 2-propanol water in the rinse solution bottle is dirty, replace with
new solution. If getting significantly dirty, replace with new solution once a day.
Even if it does not appear dirty, replace with new solution once a week.
1
Remove the rinse cap of the automatic rinsing kit from the rinse solution bottle.
2
Pour
distilled water into the rinse solution botte and clean the inside of the rinse
solution bottle using a brush, etc.
3
Pour distilled water into the rinse solution bottle and rinse the inside of the
rinse solution bottle for about 5 minutes using an ultrasonic bath.
4
Pour new 10% 2-propanol water into the rinse solution bottle, and return the
cap and rinse solution bottle back to their original positions.
LC-40D X3 CL LC-40D XR CL
123
4.16 Cleaning of the Automatic Rinsing Kit
4