5.1 Specifications
5.1.1
LC-40D X3 CL Specifications
Item Specification
Pump Type
Micro-volume double plunger pump
(approx. 10 µL/stroke)
Pumping Methods
Constant flow pumping and constant pressure
pumping
Allowable Maximum Pressure 130 MPa
Constant Flow
Pumping
Flow Rate Setting
Range
0.0001 to 10.0000 mL/min
Maximum Delivery
Pressure
130 MPa
(0.0001 to 3.0000 mL/min)
80 MPa (5.0000 mL/min)
22 MPa (10.0000 mL/min)
Flow Rate Accuracy
±1 %
(1 mL/min when water is pumped at 80 MPa,
with water and room temperature constant
between 20 to 30 °C)
Flow Rate Precision
Larger value of either 0.06 % RSD or 0.02 min
SD
Pressure Display Accuracy Larger value of either ±2 % or ±2.4 MPa
Suction Filter 10 µm
Line Filter 5 µm
Time Program
Commands for flow rate, pressure, [LOOP] (for
program repetition), 320 steps (total of 10
program files)
*1
When water is pumped at 10 to 40 MPa, with water and room temperature constant between 20
to 30 °C.
*2 When water is pumped at 40 to 60 MPa, with water and room temperature constant between 20
to 30 °C.
5
Technical Information
LC-40D X3 CL LC-40D XR CL
125
5

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Shimadzu LC-40D XR CL and is the answer not in the manual?

Shimadzu LC-40D XR CL Specifications

General IconGeneral
BrandShimadzu
ModelLC-40D XR CL
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals