5.1.2
LC-40D XR CL Specifications
Item Specification
Pump Type
Micro-volume double plunger pump
(approx. 10 µL/stroke)
Pumping Methods
Constant flow pumping and constant pressure
pumping
Allowable Maximum Pressure 70 MPa
Constant Flow
Pumping
Flow Rate Setting
Range
0.0001 to 10.0000 mL/min
Maximum Delivery
Pressure
70 MPa
(0.0001 to 3.0000 mL/min)
44 MPa
(3.0001 to 5.0000 mL/min)
22MPa
(5.0001 to 10.0000 mL/min)
Flow Rate Accuracy
Larger value of either ±1 % or ±2 µL/min
(0.01 mL/min to 3 mL/min)
Larger value of either ±2 % or ±2 µL/min
(0.01 mL/min to 3 mL/min)
Flow Rate Precision
No more than the large value calculated with
either 0.06 % RSD or 0.02 min SD
Constant Pressure
Pumping
Pressure Setting
Range
1.0 to 60 MPa
(in steps of 0.1 MPa)
Pressure Accuracy Larger value of either ±10 % or 1.5 MPa
Pressure Display Accuracy Larger value of either ±2 % or ±1.0 MPa
Suction Filter 10 µm
Line Filter 5 µm
Time Program
Commands for flow rate, pressure, [EVENT],
[LOOP] (for program repetition), 320 steps (total
of 10 program files)
126
LC-40D X3 CL LC-40D XR CL
5 Technical Information

Table of Contents