4.6 Rinsing the Check Valve
This section will explain how to pump 2-propanol to rinse the check valve.
Necessary tools
Part Name Quantity Part No.
Wrench (8 mm and 10 mm) 2 086-03002
Necessary parts
Part Name Part Type Part No. Quantity
RESISTANCE TUBE,COIL
FOR CLEANING
Replacement
part
228-57908-41 1
1
After using buffer solution as the mobile phase, replace the solvent in the
reservoir bottle with water.
After using buffer solution as the mobile phase, do NOT replace with an
organic solvent such as 2-propanol without first replacing with water.
When using a buffer solution in the reservoir bottle, replace with distilled
water or purified water, and after pumping at least 20 mL into the instrument,
replace with new 2-propanol. If the buffer solution is directly replaced with
2-propanol, crystals will be precipitated, which can cause damage to the
plunger seal or cause problems with the check valve.
System Guide "Precautions When Replacing the Mobile Phase"
2
Replace the solvent in the reservoir bottle with 2-propanol.
LC-40D X3 CL LC-40D XR CL
107
4.6 Rinsing the Check Valve