5.1.3
High-Pressure Gradient System
Item
Specification
LC-40D X3 CL LC-40D XR CL
Number of Solvents Mixed 2 or 3
Gradient Profile
Step, linear and exponential functions (Only when
controlled by LC workstation) possible at multiple levels
Maximum Program Steps 400 steps (total of 20 program files)
Program Duration 0.01 to 9999.90 minutes (in steps of 0.01 minutes)
Range of Set Concentrations 0 % to 100 % (in steps of 0.1 %)
Concentration Accuracy
*1
±0.5 %
(1 mL/min, 80 MPa)
±0.5 %
(0.05 to 3 mL/min,
1 to 40 MPa)
Flow Rate Possible 0.0001 mL/min to 10 mL/min
*1 For binary gradient with water/caffeine solution
5.1.4
Low-Pressure Gradient System
Item Specification
Number of Solvents Mixed Max. 4
Gradient Profile
Step, linear and exponential functions (Only when controlled by
LC workstation) possible at multiple levels
Maximum Program Steps 400 steps (total of 20 program files)
Program Duration 0.01 to 9999.90 minutes (in steps of 0.01 minutes)
Range of Set
Concentrations
0 % to 100 % (in steps of 0.1 %)
Concentration Accuracy
±0.5 %(1 mL/min, 10 MPa)
(for binary gradient with water/caffeine solution)
Flow Rate Possible 0.0001 mL/min to 10 mL/min
LC-40D X3 CL LC-40D XR CL
127
5.1 Specifications
5

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Shimadzu LC-40D XR CL and is the answer not in the manual?

Shimadzu LC-40D XR CL Specifications

General IconGeneral
BrandShimadzu
ModelLC-40D XR CL
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals