4
Set the compressibility.
1 Displays [COMP] of [PARAMETER]
setting group.
2 Press [].
The compressibility setting screen
is displayed.
Press [] several times to display
the [COMP B], [COMP C], and
[COMP D] setting screens.
3 Set the compressibilitys of solvents A to D.
5
Purge flow paths.
1
Open the drain valve by turning it 90 degrees counterclockwise.
2
Press [PURGE].
The flow path selection screen is
displayed.
3
Select the flow path to purge with the numeric keypad and press [Enter].
[PUMP] on the display is highlighted and purging starts.
Set Value
Description
0
Solvent is purged at the concentration set on the concentration
setting screen.
1 Solvent A is purged.
2 Solvent B is purged.
3 Solvent C is purged.
4 Solvent D is purged.
6
Close the drain valve and start delivery.
LC-40D X3 CL LC-40D XR CL
77
3.8 Delivery in the Low-Pressure Gradient Mode