① Connect to DN on flowmeter for metal pipe
(Receive) / Connect
to UP on flowmeter for
plastic pipe (Transmit)
② Connect to UP on flowmeter for metal pipe
(Receive) / Connect to DN on flowmeter for
plastic pipe (Transmit)
③ Flow Direction
Figure 7-3 In-Line Sensor Mounting
Connecting Sensor Cables
Connect sensor cables between the sensors and the flowmeter. The XMIT (transmit) sensor
is connected to the Up cable connection and RCV (receive) sensor is connected to the Down
cable connection. The UP/DN orientation is important for the in-line sensor configuration. It is
not a factor for the adjacent configuration.
Select Reflexor Operating Mode
Select Meter Type as [Reflexor] for the channel that is to be used as Reflexor. Dual channel
units may have either one or both channels used in the Reflexor mode.
Functions
7.11 Reflexor
FUP1010 IP67 Portable Flowmeter
100
Operating Instructions, 02/2010, A5E02951522A Revision 01