● If the flow rate is very high and the flowmeter is reporting erroneous or unstable flow, then
the flowmeter may already be having trouble resolving the upstream and downstream
signals. In this event, it may be necessary to first lower the flow rate to a moderate level
before performing
the "Automatic" HF adjustment. Once this is done the flowmeter should
be able to properly measure the highest flow rates without problems.
● The limits of the "HF" parameter are +/- 0.7 and any attempt to manually install a larger
value will cause the flowmeter to abort the installation of the parameter.
Note
Pressing the <Left Arrow> at any stage prior to accepting the measured value will abort
the installation and return to the previous setting.
Accessing the [HF] Function
1. At
the
[Meter
Type] Menu, press the <Right Arrow> and then <ENT> to select the desired
Path (e.g., Dual Path Flow).
2. In the [Dual Path Flow] Menu, press the <Down Arrow> and scroll to the [Diagnostic Data]
menu cell. Press the <Right Arrow> to select it.
3. In the [Diagnostic Data] Menu, highlight [Path Select] and select the desired sensor path.
Press <ENT> to select path.
4. Press the <Down Arrow> and scroll to the [Site Setup Data] menu cell. Press the <Right
Arrow> to select it.
Troubleshooting
10.4 Site Setup Data