10 Troubleshooting
196
GA-070T-0518V2.7-HAO-EN
Fritz Stephan GmbH
Alarm text
Priority/
type of
alarm
Sound
Possible cause
Corrective action
Battery 2 failure
HW
Battery 2
malfunction during
mains operation or
during battery
operation
Check battery 2, replace
if necessary
Battery 1 cal
HW
Battery 1 requires
calibration
Completely discharge
battery 1 and recharge
Battery 2 cal
HW
Battery 2 requires
calibration
Completely discharge
battery 2 and recharge
Battery 1 cycles > 500!
HW
Battery 1 has been
charged more than
500 times
Replace battery 1
Battery 2 cycles > 500!
HW
Battery 2 has been
charged more than
500 times
Replace battery 2
Alarm speaker failure
HW
Speaker malfunction
Only an acoustic
auxiliary alarm
sounds
(Piezo)
Check speaker,
replace if necessary
Contact
FRITZ STEPHAN GMBH
customer service
Blower failure 4
HW
Internal temperature
sensor malfunction
Contact
FRITZ STEPHAN GMBH
customer service
CO
2
pressure out of
range!
HW
Ambient pressure
out of range
Check ambient pressure
CO
2
temp. out of
range!
HW
Internal sensor
temperature is
outside the operating
range
Check the outside
temperature
CO
2
accuracy out of
range!
HW
The measured CO
2
concentration is
outside the specified
measuring range
Disconnect the
measuring probe and the
adapter from the tube
system, replace the
adapter, and wait 10 s.
Run zero calibration for
the CO
2
sensor.
Clean the CO
2
sensor; if
necessary, replace the
CO
2
adapter