10 Troubleshooting
Fritz Stephan GmbH
GA-070T-0518V2.7-HAO-EN
197
Flow measurement
failure!
HW
Flow measurement
outside range
Contact
FRITZ STEPHAN GMBH
customer service
Device temp. high
HW
Device temperature
high
Remove the device from
direct sunlight or other
heat source, if necessary
Check the fan operation
Check the coarse filter at
the filter inlet
Check the device
If the error occurs
despite a low ambient
temperature, contact
FRITZ STEPHAN GMBH
customer service
Change HEPA filter
HW
Service interval for
the HEPA filter
exceeded
HEPA filter exhibits
increased resistance
Change the filter
Pressure limit!
HW
The inspiration
pressure needed to
reach the preset tidal
volume is higher
than the PAW alarm
limit setting
Check the patient's
condition!
Correct the PAW alarm
limit, if necessary
Correct the target volume
setting, if necessary
Paw limited by
Pawlim
HW
Paw alarm limit
limited by Pawlim
Check Paw alarm
settings
Paw limited by Pinsp
HW
Paw alarm limit
limited by Pinsp
Check Paw alarm
settings
Paw limited by
Δ Psupp
HW
Paw limited by
Δ Psupp
Check Paw alarm
settings
Check Δ Psupp setting
Paw limited by PEEP
HW
Paw limited by
PEEP
Check Paw alarm
settings
Check PEEP setting
PEEP limited by
PEEPlim
HW
PEEP limited by
upper pressure alarm
limit
Check PEEP setting