10 Troubleshooting
204
GA-070T-0518V2.7-HAO-EN
Fritz Stephan GmbH
PEEP high
(Alarm delay: 3 s)
MP
PA
"PEEP" measured
value is above upper
PEEP alarm limit
Check the PEEP setting
and adapt to the selected
ventilation mode if
necessary
Check inspiration and
expiration times
Check the tube system
for blockages and water
droplets
Check the patient filter
and replace if necessary
Check the expiration
valve and replace if
necessary
Check CO
2
adapter
MP
TA
Incorrect CO
2
adapter
CO
2
adapter
malfunction
CO
2
adapter
contaminated
Check the CO
2
adapter
and clean if necessary
Replace the CO
2
adapter
if necessary
Standby
MP
TA
Device has been
switched to Standby
mode
No action necessary
Temp. Gasoutlet high
MP
TA
The temperature at
the blower module's
gas outlet is too high
Keep the device away
from direct sunlight or
other heat sources.
Reduce ventilation
pressures
Contact
FRITZ STEPHAN GMBH
customer service if the
temperature cannot be
reduced.
Blower failure 7!!
MP/HP
TA
If the temperature in
the blower is too
high, an MP alarm is
generated.
If the blower
overheats, an HP
alarm is triggered
Keep the device away
from direct sunlight or
other heat sources.
Reduce ventilation
frequency
Reduce flow for O
2
therapy
Battery 1 & 2 failure
HP
TA
Battery 1 and 2
malfunction
Check the batteries,
replace if necessary