10 Troubleshooting
Fritz Stephan GmbH
GA-070T-0518V2.7-HAO-EN
205
Battery 1 failure
HP
TA
Battery 1
malfunction during
mains or battery
operation
Check the battery,
replace if necessary
Battery low < 5 min
HP
TA
Battery voltage low
Ventilator has a
remaining run time
of less than 5
minutes
Check the mains supply
and charge the battery
Install the spare battery
If the battery cannot be
charged, contact
FRITZ STEPHAN GMBH
customer service
Battery 1 & 2 temp.
high
HP
TA
Both batteries have
overheated
Check the batteries,
replace if necessary
Keep the device away
from direct sunlight or
other heat sources.
Contact
FRITZ STEPHAN GMBH
customer service
Battery 1 & 2 cal
HP
TA
Both batteries
require calibration
Calibrate the batteries
Battery 1 & 2 temp.
low
HP
TA
Both batteries
undercooled
Check battery 1 and 2,
replace if necessary
Do not operate the
device at temperatures
outside ambient
conditions
(see chapter 1.9).
Contact
FRITZ STEPHAN GMBH
customer service
Airway pressure high
HP
PA
Measured
inspiration pressure
is above the upper
PAW alarm limit
Airway tube kinked
The EVE
TR
automat-
ically reduces the
pressure to PEEP level
Check the patient's
condition
Check the pressure alarm
limits
Check the flow sensor
and breathing tube
Check the ETT or mask