EasyManuals Logo
Home>Tektronix>Portable Generator>TG8000

Tektronix TG8000 Service Manual

Tektronix TG8000
372 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #118 background imageLoading...
Page #118 background image
AGL7 module adjustment procedures
10. Adjust R470 so t
hat the CH2 signal is delayed by 4 ±0.5 ns from the
CH1 signal. (See Figure 4-4.)
Figure 4-4
: Timing relationship between CH1 and CH2 s ignals
11. Move the C
H1 probe from TP501 to TP502.
12. Change the horizontal ne timing of the BLACK 3 ou tput from –7.0 ns to
+7.0 ns a
nd verify that the CH2 signal moves at least 13.5 ns.
13. If the CH2 signal moves less than 13.5 ns, turn R471 to its center position
and the
n perform the following steps:
a. Move the CH1 probe from TP502 to TP501.
b. Set the horizontal ne timing of BLACK 3 output to 0.00 ns.
c. Adjust R470 so that the CH2 signal is delayed by 4 ±0.5 ns from the
CH1 signal.
d. Change the horizontal ne timing of the BLACK 3 output from –7.0 ns to
+7.0 ns and verify that the CH2 signal moves at least 13.5 ns.
If the CH2 signal still moves less than 13.5 ns, turn R471 completely
clockwise and then repeat parts a through d.
14. Change the CH2 probe c onnection from TP430 to TP410.
15. Adjust R472 so that the CH2 signal is delayed by 8 ±0.5 ns from the
CH1 signal.
16. Change the CH1 probe c onnection from TP502 to TP504.
4–10 TG8000 Multiformat Test Signal Generator Service Manual

Table of Contents

Other manuals for Tektronix TG8000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TG8000 and is the answer not in the manual?

Tektronix TG8000 Specifications

General IconGeneral
BrandTektronix
ModelTG8000
CategoryPortable Generator
LanguageEnglish

Related product manuals