HDVG7 module adjustment procedures
HDVG7 output am
plitude adjustment table. Print this table for use during the
adjustment procedure.
Table 13-2: HDVG7 output amplitude
Minimum Value Maximum
Characterization
DMM Measure
ment
(typically 0.2880 V)
—— ——
Oscilloscope Cycle R MS
(typically 116 mV)
—— ——
Oscilloscope Cycle M ean
(typical
ly 1 mV)
—— ——
Sine wave
RMS amplitude
(typically 116 mV)
——
Attenua
tion Factor
2.35 2.55
ValueAdjustment Record
Minimum Before adjustment After adjustment Maximum
Signal 1 amplitude
Measu
red value (μ)
with attenuation
720 mV 880 mV
Black
1 amplitude
(Option BK only)
Measured value (μ)
with
attenuation
720 mV 880 mV
TG8000 Multiformat Test Signal Generator Service Manual 13–7