SDI7 module adjustment procedures
SDI output ampl
itude adjustment table. Print this table for use during the
adjustment procedure.
Table 14-2: SDI7 module SDI output amplitude
Minimum Value Maximum
Characterization
DMM Measure
ment
(typically 0.2880 V)
—— ——
Oscilloscope Cycle RMS
(typically 116 mV)
—— ——
Oscilloscope Cycle Mean
(typical
ly 1 mV)
—— ——
Sine wave
RMS am plitude
(typically 116 mV)
——
Attenua
tion Factor
2.35 2.55
ValueAdjustment Record
Minimum Before adjustment After adjustment Maximum
Signal 1A amplitude
Measu
red value (μ)
with attenuation
776 mV 824 mV
Signal 1B amplitude
Measured value (μ)
with
attenuation
776 mV 824 mV
Sign
al 2A amplitude
Measured value (μ)
with attenuation
776
mV
824
mV
Signal 2B amplitude
Me
asured value (μ)
with attenuation
776 mV 824 mV
14–6 TG8000 Multiformat Test Signal Generator Service Manual