When setting an instrument on an instrument table, pay attention
not to injure the patient's fingers between the instrument and the
table.
33
To avoid potential injury, do not touch the printer body while the
printer is in operation or when replacing the printer paper.
86
159
Before starting measurement, be sure to confirm the measuring
screen (to avoid making the ocular pressure measurement to the
patient's eye wearing a contact lens.)
101
106
119
126
Before measuring, check if there is any foreign matter on and
around the measuring nozzle.
If any, it may enter and damage the patient's eye during the mea-
surement.
110
Before measuring, set the safety stop to prevent the measuring
window glass from hitting the patient's eye.
Set it respectively for the right and left eyes.
114
Set the safety stop from the side face of the instrument.
Setting operations from other positions, where checking of the eye
position is not easy, may cause injury.
114
Clean the measuring window glass, measuring nozzle and the win-
dow glass inside the measuring nozzle using ethanol.
Other chemicals may cause damage to the patient's eye during
measurement.
156
157
This instrument has been tested (with 100/120/230V) and found to
comply with IEC60601-1-2: Ed.2.1: 2004.
This instrument may emit a radio frequency energy (within the stan-
dard) and cause an adverse effect on the devices located nearby.
When such effect is confirmed by the ON/OFF operation, a proper
treatment as to change the direction, use another outlet, etc. is rec-
ommended. If the problem remains unsolved, call your dealer.
—