9.7 Behavior of the measuring system process data outputs
State
Safety instrumented data
Non-safety instrumented data
IOPS = BAD
Data are set to 0
Data are set to 0
Connection abort
Data are set to 0
Values retain the last value before the
abort
Supply ON
Values are initialized to 0
Values are initialized to 0