SonixTouch Q+ Service Manual 00.053.204, Revision A Chapter 11: Maintenance
11-11
11.3.5 Cleaning/Disinfecting Non-Invasive Transducers
To prevent biological materials (bioburden) from drying on the equipment, always reprocess transducers as soon
as is reasonably practical following use.
Table 11-3: Non-Invasive Cleaning/Disinfecting Agents
NON-INVASIVE
T
RANSDUCERS
CLEANING/DISINFECTING AGENTS
75% IPA
Alkazyme
Cidex Activated Dialdehyde
Solution 14 day
Cidex Plus 28 day
Cidex OPA
Cidezyme
Klenzyme
McKesson Brand
Metrizyme
Milton Disinfecting Liquid
Nuclean
Omnicide – FG2
Steranios 2%
T-spray
MISCELLANEOUS
SA4-2/24
PA7-4/12
♦♦
MC9-4/12
HST15-8/20
♦♦♦♦♦
4DC7-3/40
PA4-2/20
♦♦♦♦♦
L40-8/12
m4DC7-3/40
♦♦♦♦ ♦♦
4DL14-5/38
GPS
C5-2/60 GPS
♦♦♦♦ ♦♦
L14-5/38 GPS
LINEAR AND CONVEX
C5-2/60
♦♦♦♦♦♦♦
C7-3/50
C9-5/10
♦♦♦♦♦♦♦
L9-4/38
L14-5/38
♦♦♦♦♦♦♦
L14-5W/60
Caution: Use only Ultrasonix recommended cleaners/disinfectants (Table 11-3). They have been tested and
determined safe to use on Ultrasonix transducers. Failure to follow these instructions may cause damage
and will void transducer warranties.