Chapter 7: Field Service Components 00.053.204, Revision A SonixTouch Q+ Service Manual
7-2
To Export User Data:
1. Ensure the system is plugged in and powered on.
2. Connect the external USB storage device on which the Export will be saved.
3. Tap the touch screen Menu button.
4. Select Administrator > System > Export….
5. Check the item(s) to be exported.
6. Select OK to begin the export process or Cancel to exit without exporting.
7. If OK is selected in step 6, a completion dialog will be presented when the export process has finished (this will
take approximately 15-45 seconds).
Note: When creating a backup prior to servicing or replacing the modulo, Ultrasonix recommends selecting all
options for Export.
Ensure the backup media (Device Status) has been properly detected. If the External Storage Detected
is (Q:|), the backup process will fail. In this example, the External Storage Detected is (E:\).
Use Select All to check all items and Clear All to clear all checkboxes in one step.
Only active Protocols with changes to default Presets, Measurement and Worksheet settings will be
available for Export.