Chapter 7: Field Service Components 00.053.204, Revision A SonixTouch Q+ Service Manual
7-32
3. Connect the PCI cables as below.
4. Connect the multi-colored and black power cables to connectors 3 and 4.
5. Connect the CAT5 cable to the CAT5 connector (5) on the top, right-hand side of the front block/ultrasound
module.
Note: The ribbon cables coming from the PCI card are connected in reverse. The top cable from the PCI card
(1) is connected to the bottom PCI connector on the front block/ultrasound module (1). Consequently, the
bottom cable from the PCI card (2) is connected to the top PCI connector on the front block/ultrasound
module (2).
Note: The order in which these cables are connected does not matter.