wanting to evacuate the product between linked tests or when the
pressure is so high that removing the part by hand could cause
injury.
10. Press the NEXT keypad to highlight
Reject
Level
. Reject level is the
amount of pressure drop from a stabilized pressure during test
time. Use the INCREMENT and DECREMENT keypads to set the value
for the reject level—generally a small number.
11. Press the NEXT keypad to highlight
Volume
. Volume is only used
when the Sprint LC is to convert the pressure drop to a cc/m flow
rate. If desiring test results in pressure decay, then leave this value
at zero. Press BACK to exit back to the main parameter menu.
11. Press the NEXT keypad to highlight
Next
Program
. Use the
INCREMENT and DECREMENT keypads to set the number of the next
program to link. When not using linking, set to zero.
12. Press the BACK keypad when everything is entered and ready. After
pressing BACK, Sprint LC returns to the setup menu.
Number Accelerator
While holding down the
INCREMENT
or
DECREMENT
keypad, the
ENTER/BACK
keypad
is relabeled as
ACCEL.
Tapping
ACCEL
while
holding down
INCREMENT
or
DECREMENT
will
make the values zip forward or backward.