Disable heating before performing the verifier test.
1. Slip the verifier over the 3 transducers.
2. To see the wind speed measurement, establish a terminal connection, and give the
command TABLE.
3. Make sure that WMT700 reads less than 0.22 m/s (0.5 mph) with the
verifier in place.
4. Remove the verifier.
5.7.2
Adding WMT700 to System
Before configuring the new sensor, complete the mechanical installation.
1. In AWS Client, select Settings > Parameters > Static.
2. In the CFG parameter set, select WMT700_1 or WMT700_2.
3. To enable the sensor, type 1 in the Value field, and select Save.
4. Set the wind direction oset.
a. In the General Settings parameter set, select wdoset_1 or wdoset_2.
b. In the Value field, type the wind direction oset in degrees.
c. Select Save.