190 ACF5000 FTIR ANALYZER SYSTEM | OI/ACF5000-EN REV. A
No. A
W
F S Short and long text Reaction/Comment
2530 x x Ambient temp. >>
The ambient temperature exceeds limit value
2.
For information
2531 x FTIR purging <<
The flow rate of FTIR purging falls below the
limit value.
Check settings of the pressure and flow
controller.
2534 x x Cabinet temp. >>
The internal cabinet temperature is too high.
The analyzer heaters were deactivated. To
activate the heaters, please select the menu:
Service/Test Emergency Purging
See message text
2535 Bypass activated
The air purifier is inactive. The analyzer is
purged via the bypass and not via the air puri-
fier.
For information