Reset Film Type
Follow this procedure:
Step 1: Press RESET.
Step 2: Press MENU.
Step 3: Select SET FILM TYPE from the menu. Change to a film type
other than the one currently specified. This resets the look-
up table, brightness, contrast and repeat values to different
settings, clearing the calibration error.
Step 4: If the PCR II Plus calibrates, then change back to the origi-
nal film type and recalibrate.
Reset Defaults
Caution:
This procedure resets the PCR II Plus to its default
settings and erases all custom settings. Make sure
that you have records of all custom data and look
up tables before you proceed.
Follow this procedure:
Step 1: Press RESET.
Step 2: Press MENU.
Step 3: Select RESET DEFAULTS from the menu. This resets the
look-up table, brightness, contrast and repeat values to their
factory settings, clearing the calibration error.