Battery Problems
When you turn on the film recorder, it checks look up table, color fil-
ter correction and DAC values. These values are stored in memory
powered by a back-up battery. If an error is detected, one of the these
messages appears in the message window:
?LUT CHECKSUM ERROR
The look-up table values for the current film module (and maybe
others) have been corrupted. To reload them, follow the procedure
Reloading Look Up Table Values in this section.
BATTERY! ERROR
The factory-set color filter correction and DAC values are no longer
intact. To reset them, follow the procedures Resetting Filter Correc-
tion Values in this section.
Reloading Look Up Table Values
Follow this procedure to reload look up table values:
Step 1: Press the RESET button on the PCR II Plus front panel to
clear the error message and render the unit operational.
Step 2: Next, to reload and store default brightness and contrast set-
tings, select the SET FILM TYPE function via the front panel
display.
Step 3: To customize brightness and contrast settings, follow the
procedures described in Chapter 6. If the error message ap-
pears again after power-up, call for service.
Resetting Filter Correction Values
In the event of a battery failure, enter the color correction filter values
recorded on the label on the back panel. Follow this procedure:
Step 1: Press RESET to clear the BATTERY! ERROR message.
Step
2:
Press
MENU
once,
then
press
PREVIOUS
or
NEXT
until
the
message SET BRIGHTNESS SELECT FUNCTION appears.
7-8 • PCR II Plus User Guide