EasyManua.ls Logo

AGFA PCR II Plus - Code 92 (PC): GPIB Interface Error; Code 92 (Macintosh): SCSI Interface Error; Code 93: CRT Brightness or Contrast Calibration Error

AGFA PCR II Plus
167 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Code 92 (PC): GPIB Interface Error
Secondary Code: 00
When
an
operator
attempted
to
start
a
local
test
pattern,
the
film
re-
corder was unable to put the GPIB interface into the local mode of op-
eration.
The
film
recorder
did not
initiate
a
test
pattern.
Secondary Code: 0l
The GPIB experienced a handshake error while transmitting or receiv-
ing data. Check your connections.
Code 92 (Macintosh): SCSI Interface Error
Secondary Code: 00
The film recorder has detected a parity error during data transfer from
the host.
Secondary Code: 01
SCSI Controller fault.
Code 93: CRT Brightness or
Contrast Calibration Error
Secondary Code: 00, 0l
Code 00 indicates that the CRT beam is too bright.
Code
01
indicates
that
the CRT
beam
is too
dim.
If the
film
recorder
is
unable to maintain proper CRT intensity, contact your Agfa service rep-
resentative.
Appendix A-17