Calibration Errors
The messages BRT CAL and CNT CAL indicate calibration errors. The
PCR II Plus is unable to calibrate due to one of the following conditions:
The brightness or contrast values have been modified to a level that
exceeds the CRT's internal calibration range.
A hardware failure has occurred.
Four different procedures are described in this section. In the event of
a calibration error, try them in this order:
1. Calibration Out of Range
2. Reset Film Type
3. Reset Defaults
For example, if the procedure Calibration Out of Range does not clear
the problem, go on to the next procedure on the list. If the error per-
sists
after
you try all
four
procedures,
call
for
service.
Calibration Out of Range
If you suspect that there is a problem with the internal calibration of
the PCR II Plus, initiate the calibration routine using the film recorder's
front panel keys.
Step 1: Press MENU then press NEXT or PREVIOUS until the mes-
sage window displays CRT CALIBRATION SELECT FUNC-
TION.
Step 2: Press SELECT to initiate the calibration routine, which re-
sets the calibration points to their optimal levels.
7-6 PCR II Plus User Guide