Status and Error Messages
This section shows the status and error messages and briefly explains
what they mean.
WARMING UP ROM OK RAM OK
PCR II Plus executing a power on warm-up cycle that tests its internal
electronics and calibrates its brightness levels.
CALIBRATING
The film recorder is calibrating its brightness levels.
READY AG100 PLS 4K 1
This
message
indicates
that
the
film
recorder
is
ready
to
accept
data
and shows the selected film type and resolution settings. The 1 is the
current frame count.
NO MODULE
There is no camera module attached to the film recorder. If there is in-
deed a camera module, check the connection of the camera cable and
inspect it for any bent or broken pins.
REMOTE
Indicates that the film recorder has received a command from the host
computer and that it is now under remote control.
RECORDING RED
PCR II Plus is recording the red component of a color image onto the
film.
RECORDING GREEN
PCR II Plus is recording the green color pass onto film.
RECORDING BLUE
PCR II
Plus
is
recording
the
blue
color
pass
onto
film.
END OF FILM
You have reached the end of film. Refer to the film loading and un-
loading instructions for loading fresh film and clearing the end of film
message.
7-2 • PCR II Plus User Guide