EasyManua.ls Logo

AGFA PCR II Plus - Page 117

AGFA PCR II Plus
167 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Secondary Codes: 05, 06, O7
The film recorder used the image parameter information it received and
calculated an illegal vertical video address.
• 05 indicates that the initial vertical address is greater than the al-
lowed maximum.
• 06 indicates that the calculated end address is less than the allowed
minimum.
• 07 indicates that the calculated end address is greater than the
start address.
For all three errors:
• Register pair H and L contain the vertical start address
• D and E contain the end address
• B and C contain the number of lines per scan
Secondary Code: O8
A nonexistent internal look-up table was requested.
Secondary Code: O9
The film recorder received an illegal intensity assignment.
Secondary Code: OA
The calculated horizontal endpoint was no longer within the digital
image map after the image rotated by 90 degrees.
• Register HL contains the horizontal start address.
• DE contains the horizontal end address.
• BC contains the horizontal line length.
Appendix • A-15