Detectors
Agilent 6890 Gas Chromatograph Service Manual
Auxiliary EPC 360
Replacement procedures
1. While holding the manifold by the black plastic frame as shown below,
slip the label through the slot in the mounting bracket, and align the
bracket holes over the gas fittings.
2. Finger-tighten three 7/16-inch hex nuts over the fittings to hold the
bracket in place. It is very important that you do not tighten the nuts
yet. See Figure 360-4.
Hold manifold by
the black plastic frame.
Slip label tag through
the slot in the bracket.
(FPD shown)