220 Purged/Packed Inlet
Replacement procedures
6 of 20
Jun 2001
Inlets
Agilent 6890 Gas Chromatograph Service Manual
8. Lift the inlet out of the instrument.
Figure 220-6 Removing the top inlet plate
9. If you need to remove the thermal block, slide off the insulation sleeve
and use a wrench to remove the bottom nut. Slide the thermal block off
the inlet weldment.