EasyManua.ls Logo

Agilent Technologies 7100 - Page 117

Agilent Technologies 7100
266 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Agilent 7100 Capillary Electrophoresis System User Manual 117
Using the Agilent 7100 Capillary Electrophoresis System 5
Capillary Electrophoresis, CE
The recommended maximum liquid level in the vials is 1.8 cm measured from
the bottom of the vial (see Figure 25). The minimum level depends on the type
of liquid the vial is used for (sample, buffer or waste).
Figure 25 Liquid level and lift offset for standard electrodes
Sample Vials
The level of sample in the vials depends on the injection technique used.
Please see the section “Injection” on page 103.
Liquid level
5mm
4mm
NOTE
To touch the standard electrodes (part number G7100-60007), the liquid level must comply
with the minimum condition:
Liquid Level = L = Lift offset + 5 mm up to electrode + 1 mm guard band = 1 cm. See
Figure 25.
For short electrodes (part number G7100-60033), the liquid level must comply with the
minimum condition:
Liquid Level = L = Lift offset + 10 mm up to electrode + 1 mm guard band = 1 cm.
To touch the capillary, only the liquid level must comply with the condition for both types of
electrodes:
Liquid Level = L = Lift offset + 1 mm guard band = 0.5 cm

Table of Contents

Other manuals for Agilent Technologies 7100

Related product manuals