Agilent 7100 Capillary Electrophoresis System User Manual 73
LAN Configuration 4
Manual Configuration
3 Type
?
and press [Enter] to see the available commands.
Figure 10 Telnet commands
4 To change a parameter follows the style:
• parameter value, for example:
ip 134.40.27.230
Then press [Enter]; parameter refers to the configuration parameter you are
defining, and value refers to the definitions you are assigning to that
parameter. Each parameter entry is followed by a carriage return.
Table 21 Telnet commands
Value Description
? Displays syntax and descriptions of commands
/ Displays current LAN settings
ip <x.x.x.x> Sets new IP address
sm <x.x.x.x> Sets new subnet mask
gw <x.x.x.x> Sets new default gateway
exit Exits shell and saves all changes