121
A Accessories and platepads
Installing and setting up a Vacuum Filtration Station
Bravo Platform User Guide
5 Optional. Disconnect the manual on/off toggle switch (5) from the end of the
tubing. This toggle switch is not required for the setup.
Connecting the pinch valve tubing
IMPORTANT
The house vacuum must be regulated to a maximum of 286.49 mbar (8.46 inHg),
because the pinch-valve tubing will collapse with higher vacuum pressure.
Note: The pinch-valve tubing can collapse with regular use over time, resulting in
inadequate vacuum flow for the Vacuum Filtration Station. Replace the tubing
periodically to keep the Vacuum Filtration Station functioning as intended.
The following figure shows the connections from the Vacuum Filtration Station to a
vacuum source via a pinch-valve module.
Figure Vacuum Filtration Station connections for a pinch-valve module
2
1
3
4
5
67
Item Name Description
1 Vacuum control
valve-and-gauge
assembly
Helps regulate the vacuum pressure to prevent the
pinch-valve module tubing from collapsing.
2 Pinch-valve module assembly
3-way
connector
Branches the tubing from the Vacuum Filtration Station
base to the vent and valve ports on the pinch-valve
module.

Table of Contents

Other manuals for Agilent Technologies Bravo Platform

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Bravo Platform and is the answer not in the manual?

Agilent Technologies Bravo Platform Specifications

General IconGeneral
BrandAgilent Technologies
ModelBravo Platform
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals