B Autofilling accessories
Setting up a Weigh Station
184 Bravo Platform User Guide
b In the Accessory Diagnostics dialog box, click Set Range.
The digital display shows a sensor reading that corresponds to the current
weight of the reservoir that is on the Weigh Station.
To avoid overfilling the reservoir, ensure that the full range is set while the fluid level is
below the top of the reservoir.
8 Verify that the Level reading decreases as the liquid level decreases while you run
the designated empty pump to remove any liquid from the reservoir.
9 Click OK.
10 On the Profiles tab, click Update this profile.
Related information
For information about… See…
Starting up and shutting down the
Bravo Platform
“Starting up and shutting down” on page 17
Opening Diagnostics “Opening Bravo Diagnostics” on page 24
Installing the Pump Module Pump Module User Guide
Configuring the autofilling
function
“Configuring an autofilling station for the Bravo
Platform” on page 177
Editing teachpoints “Setting teachpoints” on page 34
Using an accessory in a protocol VWorks Automation Control User Guide